Assessment and monitoring the quality of treated water and dialysate in a public hemodialysis center to prevent potential health risks

M.I. Mime, S. Bellebia, Z. Bengharez, M. Louni, K. Benrachedi

Abstract


Abstract:This study aimed to assess and monitor the physicochemical, bacteriological and endotoxin quality of dialysis water and dialysate used in a public HD center and to evaluate the performance of the unit to reject organic contaminants.

Samples were collected over two years period (October 2014 to April 2017) from different points of the hemodialysis chain. Physicochemical parameters, bacterial count and endotoxin levels in dialysis waters as well as in dialysate were measured according to international standard methods. The water treatment system used at the dialysis center is based on single reverse osmosis

The results of this study revealed the presence of chemical contaminants namely trihalomethanes (THMs), total concentration of THMs was found to be 50.69±8.980 and 42.54±19.07 μg/L in treated water and in dialysate respectively and Chloroform was the major contributor to the THMs concentration. A peak of conductivity about 4-fold higher than the norm was recorded. Detection of low microbial perturbation in both feed and treated water as well as in dialysate due probably to inappropriate disinfection.Endotoxins were identified in 84% of treated water samples with values below the limit set by the European Pharmacopoeia (<0.25 EU/mL). During the entire study period the performance of the dialysis unit for THM removal remained moderate and did not reach 50%.

More efforts are required regarding water quality management in the HD center for a successful HD therapy. Local guidelines on HD water quality should be developed and enforced.


Full Text:

PDF

References


Barsoum,R.S. Chronic kidney disease in the developing world. N Engl J Med,2006, 354(10):997–

, doi: 10.1056/NEJMp058318.

Bamgboye, E.L. Hemodialysis: Management problems in developing countries, with Nigeria as asurrogate. Kidney Int, 2000,63(Suppl 83): S93-S95. doi: 10.1046/j.1523-1755.63.s83.19.x

Shahryari, A.; Nikaeen, M.; Hatamzadeh,M.;VahidDastjerdi, M.;Hassanzadeh, A. Evaluation ofBacteriological and Chemical Quality of Dialysis Water and Fluid in Isfahan, Central Iran. Iran J Public Health, 2016, 45(5):650-656.

Marouane, J.;Fadili, W. ;Kennou, B. ;Labaali, A. ;Zahlane, K ;Laouad, I. Évaluation de l’impact d’un centre d’hémodialyse sur l’environnement et l’écologie locale. Néphrologie & Thérapeutique, 2013, 9:481–485. https://doi: 10.1016/j.nephro.2013.07.369.

European Pharmacopoeia.8th Edition. Concentrated solutions for hemodialysis. 2014.

Association for the Advancement of Medical Instrumentation. Guidance for the preparation and quality management of fluids for hemodialysis and related therapies.ANSI/AAMI/ISO 23500.2011. Arlington (VA): AAMI, 2011.

International Organization for Standardization: Guidance for the Preparation and Quality Management of Fluids for Haemodialysis and Related Therapies, ISO 23500:2011. Geneva: International Organization for Standardization, 2011.

ISO 9308-1:2000 .Water Quality-Detection and Enumeration of Escherichia coli and Coliform Bacteria-Part 1: Membrane Filtration Method. Geneva: International Organization for Standardization.

ISO 6222:1999. Water Quality-Enumeration of Culturable Micro-organisms- Colony Count by Inoculation in a Nutrient Agar Culture Medium.International Organization for Standardization.

ISO 9308-1:2000.Water Quality-Detection and Enumeration of Intestinal Enterococci –Part 2: Membrane Filtration Method. Geneva: International Organization for Standardization.

Honore, P. ; Joannes-Boyau, O. ; Collin,V. ; Boer, W. ; Gressens, B. ; Janvier, G. Gestion pratique de l’épurationextra rénale continue au quotidien. Réanimation, 2008, 17(5):472-477.

doi.org/10.1016/j.reaurg.2008.01.008

Gisselbaek, M. ;Pavlovic, G. ;Pugin, J. Traitement du choc septique par épuration extracorporelle de l’endotoxine circulante. Revue Médicale Suisse, 2013, 9 : 2330-2334.

Géhin, D. ; Le Bacle, C. Endotoxines en milieu de travail. I. Origine et propriétés toxiques des endotoxines. Métrologie. Documents pour le médecin du travail, 2011, 126 :225-240, TC 136.

Singh, A.K.; Magee, C.C.; Tucker, J.K. Core Concepts in Dialysis and Continuous Therapies. Chapter 2:Technical Aspects of hemodialysis. Springer Science+Business. Media, LLC. 2016; 292.

DOI 10.1007/978-1-4899-7657-4_2.

Formica, M.;Vallero, A.;Forneris, G.; Cesano, G.; Pozzato, M.;Borca, M.et al.Organohalogenated contamination of a dialysis-water treatment plant. G ItalNefrol, 2002, 19: 479–482.

Poli, D.;Pavone, L. ;Tansinda, P. ; Goldoni, M. ;Tagliavini, D. ; David, S. ;et al. Organic contamination in dialysis water: trichloroethylene as amodel compound. Nephrol Dial Transplant , 2006, 21: 1618–1625. Doi:10.1093/ndt/gfl036.

Viti, A. ; Tinti, P. ; Verrucoli, G. ; Lorini, C. ; Sacco, C. ; Lanciotti, E. Contamination by disinfection byproducts (DPBs) and microbiological quality of dialysis water: a study in an Italian hospital dialysis service. IgSanitaPubbl, 2006, 62(2) :173-187.

Mime, M.I. ; Bouterfas, B. ; Bellebia, S. ; Bengharez, Z. ; Benrachedi, K. Investigation Expérimentale et Statistique de la Qualité de L'eau pour Hémodialyse et du Dialysat. 4ème ConférenceInternationale des Energies Renouvelables (CIER-2016). Proceedings of Engineering and Technology – PET. 16:97 - 103

World Health Organization. Trihalomethanes in drinking water.Background document for development of WHO Guidelines for Drinking-water Quality.WHO/SDE/WSH/05.08/64.

AFSSA: Agence Française de Sécurité Sanitaire des Aliments: Evaluation des risques sanitaires liés aux situations de dépassement de la limite de qualité du paramètre « trihalométhanes totaux » dans les eaux destinées à la consommation humaine, 15 Avril 2010.

Bove, G.E.; Rogerson, P.A. ; Vena, J.E. Case control study of the geographic variability of exposure to

disinfectant byproducts and risk for rectal cancer. Int J Health Geogr, 2007, 6:18.doi:10.1186/1476-072X-6-18.

Villanueva, C.M. ; Gracia-Lavedan, E. ; Bosetti, C. et al. Colorectal Cancer and Long-Term Exposure toTrihalomethanes in Drinking Water: A Multicenter Case–Control Study in Spain and Italy. Environ Health Perspect, 2017, 125(1): 56-65. doi:10.1289/EHP155

WHO (2007) Desalination for safe water supply: Guidance for the health and environmental aspects applicable to desalination. World Health Organization, Geneva. Available at: http://www.who.int/water_sanitation_health/gdwqrevision/ desalination.pdf

Ziolkowska,J.R.. Is Desalination affordable? – Regional cost and price analysis. Water Resour Manage,2015, 29: 1385–1397. https://doi.org/10.1007/s11269-014-0901-y.

Coulliette,A.D.; Arduino, M.J. Hemodialysis and Water Quality. Semin Dial, 2013, 26(4): 427–438.

Doi:10.1111/sdi.12113.

Sartori, F.G .;Leandro, L.F. ;Montanari, L.B. et al. Isolation and Identification of Environmental

Mycobacteria in the Waters of a Hemodialysis Center.CurrMicrobiol, 2013, 67:107–111. doi:10.1007/s00284-013-0341-6.

Schiavano, G.F.; Parlani, L.;Sisti, M.; Sebastianelli, G.; Brandi, G. Occurrence of fungi in dialysis water anddialysate from eight haemodialysis units in central Italy. J HospInfec, 2014, 86:194-200. DOI:10.1016/j.jhin.2013.11.010

Glorieux, G. ;Schepers, E. ; Schindler, R. ;Lemke, H.D. ;Verbeke, F.;Dhondt, A.; et al. A novel bio-assay increases the detection yield ofmicrobiological impurity of dialysis fluid, in comparison to the LAL-test. Nephrol. Dial.Transplant, 2009, 24:548–554. doi: 10.1093/ndt/gfn485

Totaro, M. ;Casini, B. ; Valentini, P. ;Miccoli, M. ; Giorgi, S. ;Porretta, A. ; et al. Evaluation and control of microbial and chemical contaminationin dialysis water plants of Italian nephrology wards. J Hosp Infect, 2017, 97(2):169-174. doi:10.1016/j.jhin.2017.05.011.

Braimoh, R.W.;Mabayoje, M.O.; Amira, C.O.; Bello, B.T. Microbial quality of hemodialysis water, a surveyof six centers in Lagos, Nigeria. HemodialInt, 2014, 18(1): 148-152. doi: 10.1111/hdi.12070.

Thomé, F.S. ;Senger, M. ;Garcez, C. ;Garcez, J. ;Chemello, C. ;Manfro, R.C. Dialysis water treated by reverse osmosis decreases the levels of C-reactive protein in uremic patients. Braz J Med Biol Res, 2005, 38(5):789-794.

Oumokhtar, B.;Lalami, A.el O.; Mahmoud, M.;Berrada, S.;Arrayhani, M.;Houssaini, T.S. Prevent infectionlinked to the dialysis water in a hemodialysis center in Fez city (Morocco). Pan Afr Med J, 2013, 16:122. doi: 10.11604/pamj.2013.16.122.2877.

Kasparek, T.; Rodriguez, O.E. What Medical Directors Need to Know about Dialysis Facility Water Management.Clin J Am SocNephrol, 2015, 10(6):1061-1071.doi: 10.2215/CJN.11851214

Hasegawa, T.;Nakai, S.;Masakane, I.; Watanabe, Y.; Iseki, K.; Tsubakihara, Y.; et al. Dialysis fluid endotoxin level and mortality in maintenancehemodialysis: a nationwide cohort study. Am J Kidney Dis, 2015, 65(6): 899-904. doi:10.1053/j.ajkd.2014.12.009.